Abstract:Determination of trace elements Cd in ilmenite deposit by using graphite furnace atomic absorption method has been studied, and the best instrument condition under the matrix interference with high content of Fe, Ti and NH4H2PO4-H2NCSNH2-EDTA mixed matrix modifier has been studied. In picking sampling 0.2000g, 25 mL volumetric conditions, the detection limit was 0.006μg / g, and using a national standard to verify, its accuracy (RE%) <10%, 12 times measured precision degree (RSD%) <10%, and the recovery is 90%~110%.