Abstract:By using HCLH2NO3HFH2SO4 to melt ore, using Yinternal standard by inductively coupled plasmaatomic emission spectrometry(ICPAES) method, the method for determining Cu,Co,Ni,V and Cr in ilmenite and vanadium and titanium magnetite at the same time has been set up. The amount of fluxes,detecting lines and instrument working parameters have been selected, the matrix effects of titanium and iron, and the method for eliminating the influence of matrix by using Yinternal standard have been studied in this paper. It is showed that Yinternal standard can eliminate the influence of matrix on the detection, and the accuracy and precision of analysis results can be improved as well. Compared with classic spectrophotometry and atomic absorption spectrometry, this method has the advanges of lower detection limits, higher sensitivity, simpler and faster operation. Tested by National Standard Reference Materials, the determination values are in good agreement with certified values .